5700A/5720A Series II Calibrator
Service Manual
2-106
2-166. Calibration of the Current Function.
Refer to Figure 2-29 for the following discussion. The resistor network Z5 determines the
accuracy of the 2.2A current range. Calibration involves determining its offset and gain
constants. The Calibrator is configured as in the 2.2A dc function, except that the Current
assembly (A7) is set to the 2.2 mA range.
To determine offset, the output (B CUR) is routed back to the Current assembly where it
is connected to INT OUT HI. INT OUT HI is routed to the Ohms Main assembly (A10),
where it is connected to a previously calibrated 10 resistor. The voltage generated
across this 10 resistor is routed to the differential amplifier on the Ohms Cal assembly
(A9). The output of the differential amplifier is routed to the DAC assembly (A11) where
it is measured by its adc circuit. A checkpoint reading is first taken by removing INT
OUT HI from the 10 resistor and measuring the voltage across the resistor. The offset is
then measured by connecting INT OUT HI to the 10 resistor. The DAC assembly,
which controls the output of the Current assembly, is adjusted until the adc circuit
measures the same as the previous checkpoint reading.
To determine gain, the Calibrator is configured as in the previous step, except with the
Current assembly outputting 1.3 mA generated from the 13V reference (BRF13 and
BSRF13) from the DAC assembly. This 1.3 mA is amplified 100 times by the 2.2A
amplifier. The resulting 130 mA is connected to the 10 resistor on the Ohms Main
assembly by the same path as for the offset calibration. The resulting 1.3V across this
10 resistor is routed to the Ohms Cal assembly (A9), where it is connected to the -input
of the differential amplifier. The +input of the differential amplifier is connected to the
output of the DAC assembly. The output of the differential amplifier is connected to the
RCL line which is routed to the adc circuit on the DAC assembly. The DAC output,
which is the +input of the differential amplifier, is adjusted until a null is measured on the
RCL line by the DAC's adc circuit. This step is repeated by changing the -input of the
differential amplifier to the other side of the 10 resistor. The software now computes
the exact voltage drop across the 10 resistor. Gain is determined by using this and the
previous offset reading.
2-167. High Voltage Magnitude Control
The square wave (HVCL) used in the previously described functions, is created and
amplitude controlled by the High Voltage Control assembly (A14). This circuitry, shown
on sheet 2 of the schematic, contains the absolute value circuit, signal/polarity selection
circuit, reference and error amplifier, square wave generator, and the square wave
amplifier.
The absolute value circuit contains op amp U2A, U2B, Q3, diodes CR8 and CR9,
capacitor C20, and resistors R27-R32 and R68. During operation in the ac current
function, this circuit creates an absolute value of the G OUT signal from the collectors of
the 2.2A output transistors. Op amp U2A generates a negative half-wave signal equal to
the positive peaks of G OUT. Resistors R32 and R27 sum this half-wave signal and the
input signal G OUT at the input of U2B. Capacitor C20 averages the voltage so the
output of U2B is a dc voltage which represents the positive peak voltage of G OUT. In
the high voltage dc function, the 2.2A amplifier circuit is not used, so the output of U2B
is 0V.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series and is the answer not in the manual?

Fluke 5700A Series Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Humidity10% to 90% non-condensing
Frequency Range10 Hz to 1 MHz
AccuracyVaries by function and range. See manufacturer's specifications for details.
Output StabilityVaries by function and range. See manufacturer's specifications for details.
Power Requirements50 Hz to 60 Hz