5700A/5720A Series II Calibrator
Service Manual
1-18
1-31. AC Voltage Specifications
5720A Series II AC Voltage Specifications: 99% Confidence Level
Absolute Uncertainty
± 5 °C from calibration temperature
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Range Resolution
Frequency
(Hz)
± (ppm output + µV)
2.2 mV 1 nV 10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 5
100 + 5
85 + 5
220 + 5
500 + 6
1000 + 12
1400 + 25
2900 + 25
270 + 5
105 + 5
90 + 5
230 + 5
540 + 6
1200 + 12
1500 + 25
3100 + 25
290 + 5
110 + 5
95 + 5
240 + 5
570 + 6
1250 + 12
1600 + 25
3250 + 25
300 + 5
115 + 5
100 + 5
250 + 5
600 + 6
1300 + 12
1700 + 25
3400 + 25
250 + 5
100 + 5
60 + 5
85 + 5
200 + 6
350 + 12
800 + 25
2700 + 25
270 + 5
105 + 5
65 + 5
95 + 5
220 + 6
400 + 12
1000 + 25
3000 + 25
22 mV 10 nV 10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 5
100 + 5
85 + 5
220 + 5
500 + 6
1000 + 12
1400 + 25
2900 + 25
270 + 5
105 + 5
90 + 5
230 + 5
540 + 6
1200 + 12
1500 + 25
3100 + 25
290 + 5
110 + 5
95 + 5
240 + 5
570 + 6
1250 + 12
1600 + 25
3250 + 25
300 + 5
115 + 5
100 + 5
250 + 5
600 + 6
1300 + 12
1700 + 25
3400 + 25
250 + 5
100 + 5
60 + 5
85 + 5
200 + 6
350 + 12
800 + 25
2700 + 25
270 + 5
105 + 5
65 + 5
95 + 5
220 + 6
400 + 12
1000 + 25
3000 + 25
220 mV 100 nV 10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 15
100 + 8
85 + 8
220 + 8
500 + 20
850 + 25
1400 + 30
2700 + 60
270 + 15
105 + 8
90 + 8
230 + 8
540 + 20
900 + 25
1500 + 30
2900 + 60
290 + 15
110 + 8
95 + 8
240 + 8
570 + 20
1000 + 25
1600 + 30
3100 + 60
300 + 15
115 + 8
100 + 8
250 + 8
600 + 20
1100 + 25
1700 + 30
3300 + 60
250 + 15
100 + 8
60 + 8
85 + 8
200 + 20
350 + 25
800 + 30
2600 + 60
270 + 15
105 + 8
65 + 8
95 + 8
220 + 20
400 + 25
1000 + 30
2800 + 60
2.2 V 1 µV 10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 50
95 + 20
45 + 10
80 + 12
120 + 40
380 + 100
1000 + 250
1600 + 400
270 + 50
100 + 20
47 + 10
85 + 12
125 + 40
420 + 100
1100 + 250
1800 + 600
290 + 50
105 + 20
50 + 10
87 + 12
127 + 40
460 + 100
1150 + 250
1900 + 400
300 + 50
110 + 20
52 + 10
90 + 12
130 + 40
500 + 100
1200 + 250
2000 + 400
250 + 50
95 + 20
30 + 10
70 + 12
100 + 40
270 + 100
900 + 250
1200 + 400
270 + 50
100 + 20
40 + 10
75 + 12
105 + 40
290 + 100
1000 + 250
1300 + 400
22 V 10 µV 10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 500
95 + 200
45 + 70
80 + 120
110 + 250
300 + 800
1000 + 2500
1500 + 4000
270 + 500
100 + 200
47 + 70
85 + 120
115 + 250
310 + 800
1100 + 2500
1600 + 4000
290 + 500
105 + 200
50 + 70
87 + 120
117 + 250
320 + 800
1150 + 2500
1700 + 4000
300 + 500
110 + 200
52 + 70
90 + 120
120 + 250
325 + 800
1200 + 2500
1800 + 4000
250 + 500
95 + 200
30 + 70
70 + 120
100 + 250
270 + 800
900 + 2500
1300 + 4000
270 + 500
100 + 200
40 + 70
75 + 120
105 + 250
290 + 800
1000 + 2500
1400 + 4000
± (ppm output + mV)
220 V
[2]
100 µV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
250 + 5
95 + 2
57 + 0.7
90 + 1.2
160 + 3
900 + 20
5000 + 50
8000 + 100
270 + 5
100 + 2
60 + 0.7
95 + 1.2
170 + 3
1000 + 20
5200 + 50
9000 + 100
290 + 5
105 + 2
62 + 0.7
97 + 1.2
175 + 3
1050 + 20
5300 + 50
9500 + 100
300 + 5
110 + 2
65 + 0.7
100 + 1.2
180 + 3
1100 + 20
5400 + 50
10,000 + 100
250 + 5
95 + 2
45 + 0.7
75 + 1.2
140 + 3
600 + 20
4500 + 50
8000 + 100
270 + 5
100 + 2
50 + 0.7
80 + 1.2
150 + 3
700 + 20
4700 + 50
8500 + 100
1100 V
[1]
1 µV
15 - 50 300 + 20 320 + 20 340 + 20 360 + 20 300 + 20 320 + 20

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series and is the answer not in the manual?

Fluke 5700A Series Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Humidity10% to 90% non-condensing
Frequency Range10 Hz to 1 MHz
AccuracyVaries by function and range. See manufacturer's specifications for details.
Output StabilityVaries by function and range. See manufacturer's specifications for details.
Power Requirements50 Hz to 60 Hz