EasyManua.ls Logo

Fluke 5700A Series - Waveform at TP4, Calibrator Set to 6.5 V; Waveform at TP7

Fluke 5700A Series
576 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
Component-level Troubleshooting 5
5-59
1 V
10 ms
F5-19.EPS
Figure 5-19. Waveform at TP4, Calibrator Set to 6.5V
5. Check CH2 FILTER INPUT. Connect an oscilloscope probe to TP7 (common to
TP3). Set the Calibrator to 6.5V dc, operate. Set the oscilloscope to 2V/div at 2
ms/div. Verify the oscilloscope displays pulses similar to those shown in Figure 5-20.
1 V
10 ms
F5-20.EPS
Figure 5-20. Waveform at TP7

Table of Contents

Related product manuals