EasyManuals Logo

Fluke 5700A Series User Manual

Fluke 5700A Series
576 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #347 background imageLoading...
Page #347 background image
Troubleshooting
Component-level Troubleshooting 5
5-51
5-8. Hi-Res Oscillator Section
Proceed as follows to troubleshoot the Hi-Res Oscillator Section of the Current/Hi-Res
assembly (A7):
1. Remove the four small shields (two on each side) and place the Current/Hi-Res
assembly on the analog reverse extender card. Power up the Calibrator.
Note
All measurements are referenced to LHCOM (TP18).
2. Check the 8 MHz reference. Set the Calibrator to 2V at 1 kHz, operate. Connect an
oscilloscope to U13 pin 4 and verify it displays a 8 Mhz signal similar to that shown
in Figure 5-12. If a failure is detected, first verify that control line HI-RES ON/OFF
is a logic low by measuring U15 pin 12. If not U7 is probably at fault. If this control
line is correct check U13 and its associated components.
1 V
10 ms
F5-12.EPS
Figure 5-12. Waveform at Pin 4 of U13
Note
The input CLK and CLK* is a low level (400 mV p-p) signal generated by
the Regulator/Guard Crossing assembly.
3. Check the 2 MHZ REF. Set the Calibrator to 2V at 1 kHz, operate. Connect an
oscilloscope to TP12 and verify it displays a 2 MHz signal similar to that shown in
Figure 5-13. If a failure is detected, check U14A and U14B.
4. Check the 6-12 MHZ signal. Set the Calibrator to 2V at 1 kHz, operate. Connect an
oscilloscope to TP13 and verify it displays a 10 Mhz signal similar to that shown in
Figure 5-14. Also measure the frequency of this signal using a frequency counter and
verify it is within 9.999 MHz to 10.001 MHz. If a failure is detected, with either test
continue on as follows. If the signal and frequency are correct, skip to step 9.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series and is the answer not in the manual?

Fluke 5700A Series Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Humidity10% to 90% non-condensing
Frequency Range10 Hz to 1 MHz
AccuracyVaries by function and range. See manufacturer's specifications for details.
Output StabilityVaries by function and range. See manufacturer's specifications for details.
Power Requirements50 Hz to 60 Hz

Related product manuals