Table 3-16. 5720A Resistance Test Record
1 Std Res
Value
1 Std
Voltage
Measured
1 Std
Voltage
Cal
Current
UUT 1
Displayed
Value
UUT 1
Voltage
UUT
True
Res
UUT
Dev.
Uncert
Std
Spec
(±ppm)
TUR GB
Factor
Test
Limit
(±ppm)
Max
Difference of
Charactized to
Nominal Value
5.9 95 16.06 1.000 95 0.9995 to 1.0005
1.9 Std
Res Value
1.9 Std
Voltage
Measured
1.9 Std
Voltage
Cal
Current
UUT 1.9
Displayed
Value
UUT
1.9
Voltage
UUT
True
Res
UUT
Dev.
Spec
(±ppm)
Max
Difference of
Characterized to
Nominal Value
5.9 95 16.06 1.000 95 1.89905 to
1.90095
10 Std
Res Value
10 Std
Voltage
Measured
10 Std
Voltage
Cal
Current
UUT 10
Displayed
Value
UUT
10
Voltage
UUT
True
Res
UUT
Dev.
Spec
(±ppm)
Max
Difference of
Characterized to
Nominal Value
5.9 25 4.23 1.000 25 9.997 to 10.004
Resistance Accuracy Verification (19 And Above)
Std
Res
Value
DMM Std
Res Rdg
DMM
Error
DMM
UUT Res
Value
UUT True
Res Value
UUT Dev. Uncert
Std
Spec
(±ppm)
TUR GB
Factor
Test
Limit
(±ppm)
Max
Difference of
Characterized to
Nominal Value
19 (1) 25 18.9943 to 19.0057
100 (1) 5.1 11 2.16 0.970 10.7 99.985 to 100.015
190 (1) 11 189.9715 to 190.0285
1 k (1) 4.7 9 1.91 0.926 8.3 999.85 to 1000.15
1.9 k (1) 9 1.899715k to 1.900285k
10 k 3.1 9 2.92 0.978 8.8 9.9985k to 10.0015k
19 k 3.5 9 2.55 0.963 8.7 18.99715k to 19.00285k
100 k(1) 5.1 11 2.15 0.942 10.4 99.985k to 100.015k
190 k (1) 11 189.9715k to 190.0285k
1 M (1) 8.1 18 2.23 0.948 17.1 0.9998M to 1.0002M

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series and is the answer not in the manual?

Fluke 5700A Series Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Humidity10% to 90% non-condensing
Frequency Range10 Hz to 1 MHz
AccuracyVaries by function and range. See manufacturer's specifications for details.
Output StabilityVaries by function and range. See manufacturer's specifications for details.
Power Requirements50 Hz to 60 Hz