Introduction and Specifications
Electrical Specifications 1
1-31
Resistance Secondary Performance Specifications and Operating Characteristics
Temperature Coefficient
Adder
[2]
Two-Wire Adder
Active
Compensation
[4]
Maximum
Difference
of
Characterized
to Nominal
Value
Lead Resistance
Stability
±1 °C
[1]
24 Hours
10 - 40 °C
0 - 10 °C
and
40 - 50 °C
0.1 1
Nominal
Value
(Ω)
±ppm ±ppm/°C
Full Spec Load
Range
[3]
I
L
- I
U
(mA)
Maximum
Peak
Current
I
MAX
(mA)
±ppm
±m
0
8 - 500 500
2 +
m
I
µV 4
4 +
m
I
µV 4
1 32 4 5 8 - 100 700 500
2 +
m
I
µV 4
4 +
m
I
µV 4
1.9 25 6 7 8 - 100 500 500
2 +
m
I
µV 4
4 +
m
I
µV 4
10 5 2 3 8 - 11 220 300
2 +
m
I
µV 4
4 +
m
I
µV 4
19 4 2 3 8 - 11 160 300
2 +
m
I
µV 4
4 +
m
I
µV 4
100 2 2 3 8 - 11 70 150
2 +
m
I
µV 4
4 +
m
I
µV 4
190 2 2 3 8 - 11 50 150
2 +
m
I
µV 4
4 +
m
I
µV 4
1 k 2 2 3 1 - 2 22 150 10 15
1.9 k 2 2 3 1 - 1.5 16 150 10 15
10 k 2 2 3
100 - 500 µA
7 150 50 60
19 k 2 2 3
50 - 250 µA
5 150 100 120
100 k 2 2 3
10 - 100 µA
1 150 I
m
= Current produced
by Ohmmeter (A)
190 k 2 2 3
5 - 50 µA 500 µA
150
1 M 2.5 2.5 6
5 - 20 µA 100 µA
200
1.9 M 3.5 3 10
2.5 - 10 µA 50 µA
200
10 M 10 5 20
0.5 - 2 µA 10 µA
300
19 M 20 8 40
0.25 - 1 µA 5 µA
300
100 M 50 12 100 50 - 200 nA
1 µA
500
Notes:
1. Stability specifications are included in the Absolute Uncertainty values in the primary specification tables.
2. Temperature coefficient is an adder to uncertainty specifications that does not apply unless operated more than 5 °C from calibration
temperature, or calibrated outside the range 19 °C to 24 °C. Two examples:
- Calibrate at 20 °C: Temperature coefficient adder is not required unless operated below 15 °C or above 25 °C.
- Calibrate at 26 °C: Add 2 °C temperature coefficient adder. Additional temperature coefficient adder is not required unless operated
below 21 °C or above 31 °C.
3. Refer to current derating factors table for loads outside of this range.
4. Active two-wire compensation may be selected for values less than 100 k, with either the front panel or the meter input terminals as
reference plane. Active compensation is limited to 11 mA load, and to 2 V burden. Two-wire compensation can be used only with -
meters that source continuous (not pulsed) dc current.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series and is the answer not in the manual?

Fluke 5700A Series Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Humidity10% to 90% non-condensing
Frequency Range10 Hz to 1 MHz
AccuracyVaries by function and range. See manufacturer's specifications for details.
Output StabilityVaries by function and range. See manufacturer's specifications for details.
Power Requirements50 Hz to 60 Hz