Introduction and Specifications
Specifications
1
1-7
Table 1-2. 8062A Specifications (cont)
Resistance
Ranges...................................... 200Ω, 2 kΩ, 20 kΩ, 200 kΩ, autoranging
MΩ. The MΩ range extends from .0001
mΩ to 300 MΩ in three autoranged
ranges. Upscale range changes are
made at 2 MΩ and 20 MΩ. Downscale
range changes are made at 19 MΩ and
1.9 MΩ.
Accuracy ................................... ±(% of reading + no. of digits). See table
below.
Range
Reso-
lution
Accuracy
Full-
scale
Voltage
Max
Current
Open
Circuit
Voltage
200Ω 0.01Ω (0.1%+2+.02Ω) <1.1 mA <4.8V
2 kΩ 0.1Ω (0.1%+2)
<250 mV
<150 µA
20 kΩ 1Ω (0.1%+2) <15 µA <1.5V
200 kΩ 10Ω (0.1%+2) <1.5 µA
0-1.9999 MΩ 100Ω (0.2%+2)
MΩ
2-19.99 M
Ω 10 kΩ (0.25%+3) <2.5 µA
20-99.9 MΩ 100 kΩ (1%+3) <2.5V <2.5V
100-300 MΩ 1 MΩ (2%+3)
Autoranging kΩ
0.1Ω
to 1 kΩ
(.20%+5) <1.0 mA
Response Time.............................. Two seconds maximum to rated
accuracy for all ranges except MΩ.
For MΩ, 8 seconds maximum.
Overload Protection ....................... 300V dc or rms ac for all ranges