8062A
Instruction Manual
1-8
Table 1-2. 8062A Specifications (cont)
Continuity
Ranges........................................... All resistance ranges
Threshold....................................... Nominally <50% of range (for
example, 100 in the 200 range)
for 200, 2 k, 20 k, 200 k
ranges. Nominally <100 k in
autoranging k.
Display Indication........................... Horizontal bar across the top of the
display and/or 2.667 kHz tone.
Indication is present for a minimum of
200 ms.
Response Time.............................. 50 µs maximum (10 µs typical)
Overload Protection ....................... 300V dc or rms ac
Diode Test
Range ............................................ 2V
Test Current................................... 1 mA (typical)
Accuracy ........................................ ±(0.06% of reading + 2 digits)
(Specification applies for voltage
measurement)
Response Time.............................. 1 seconds maximum
Overload Protection ....................... 300V dc or rms ac
DC Current
Range Resolution
Accuracy
±(% of reading + no. of digits)
Burden
Voltage
200 µA
2 mA
20 mA
.01 µA
.1 µA
1 µA
0.3% + 2
.3V typical
.3V typical
.3V typical
200 mA
2000 mA
10 µA
100 µA
0.7% + 2
.3V typical
.9V typical
Overload Protection ....................... 2A/250V fuse (operator replaceable)
in series with 3A/600V fuse (service
personnel replaceable).