Introduction and Specifications
Specifications
1
1-9
Table 1-2. 8062A Specifications (cont)
AC Current (True RMS Responding, AC-Coupled
Accuracy *...................................... ±(% of reading + no. of digits). See
table below.
Input
Current
Resolution Range 20 Hz -
45 Hz
45 Hz -
3 kHz
3 kHz -
10 kHz
10 kHz -
30 kHz
20.00 to
199.99 µA
0.01 µA 200 µA
.2000 to
1.9999 mA
0.1 µA 2 mA 2% + 40
2.000 to
19.999 mA
1 µA 20 mA
1% + 10 0.75% + 10 2% + 20
20.00 to
199.99 mA
10 µA 200 mA
1.5% + 10 1% +10
200.0200 to
1999.9 mA
100 µA 2000 mA
Not Specified
* Not specified for input < 10% of scale.
Burden Voltage.............................. 0.3V rms typical except 2000 mA
range, 0.9V rms typical
Overload Protection ....................... 2A/250V fuse (operator replaceable)
in series with 3A/600V fuse (service
personnel replaceable).
Relative
Selection........................................ When the REL button is pushed, the
input applied at that time is stored as
a zero reference point. Subsequent
readings indicate deviations (±) from
the reference point.
Accuracy ........................................ Error does not exceed the sum of the
errors of the reference reading and
the subsequent reading.