EasyManua.ls Logo

Fox Thermal FT3 - Calibration Validation Troubleshooting

Default Icon
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting | 119
Model FT3
Troubleshooting
Troubleshooting CAL-V
If the FT3 meter fails a CAL-V™ Calibration Validation test, follow the steps below:
1. Flow rate in the pipe:
Run the test again under a higher flow rate if possible.
2. The sensor may be dirty or damaged:
Visually inspect the meter for damage. If damage is found, meter may need to be serviced.
Contact Fox Thermal Technical Assistance at 831-384-4300 for more information.
Try cleaning the sensor and try the test again under higher gas flow conditions.
3. If the meter continues to fail, contact Fox Thermal Technical Assistance at 831-384-4300 for
more information.
Troubleshooting Zero CAL-CHECK
If the FT3 Meter fails a Zero CAL-CHECK
Calibration Validation test, there are a few reasons that
could be the cause:
1. The sensor may be dirty or damaged
Visually inspect the meter for damage. If damage is found, meter may need to be serviced.
Contact Fox Thermal technical assistance at 831-384-4300 for more information
Try cleaning the sensor and try the test again
If the meter fails again, move to #2
2. The sensor may not be properly covered/isolated
Out-of-pipe:
Wind currents (fans in room included) could be affecting the sensor
For best results, be sure to use the factory-supplied PVC sensor cover (see Fig. 4.5 on
Page 79)
If the factory-supplied PVC sensor cover is unavailable, use a clean dry plastic beverage
bottle
In Pipe:
If using the retractor, be sure that the shut-off valve has been closed
Make sure that there is a "no flow" or zero flow condition on the meter's sensor
Try the test again
If the meter fails again, move to #3
3. The meter may not have stabilized properly
Make sure the meter is not being affected by vibrations or other movement
Allow the meter to stabilize without being moved or touched for 15 minutes
Try the test again
If the meter continues to fail, contact Fox Thermal technical assistance at 831-384-4300

Other manuals for Fox Thermal FT3