Alarm
Code
Reason Action
37 Totalizer Error
Detected
See "Reset the Total and Elapsed Time" on p. 74 for steps to clear
Error Code. Contact Fox Thermal for possible causes.
38 CAL-V™ in
progress
Wait until the CAL-V™ or Cal Set is nished.
39 ADC12 versus
ADC24 too far
The tolerance between the 2 analog to digital converters is out of
specication. Recalling manufacture default parameters using the
Restore Database command on p. 11 may correct the problem.
40 CAL-V™ Di Fail The CAL-V™ Di Failed. Check sensor wiring and verify that the
sensor's resistance is correct. Call Tech Support.
41 Zero CAL-CHECK™
Fail
Allow meter to stabilize for 15 minutes and perform the test again. If
another "Fail" test results, call Tech Support.