Write Single Register (Command 06)
This command is used to perform miscellaneous functions such as clearing the totalizer and initiating
diagnostic operations. The register address is 0x0a (10 decimal, Modbus=40011) and the data to write
is described below.
Request:
<Meter Address> <Command code=06> <Register address high=0x00> <Register address
low=0x0a> <Register data high=0x00> <Register data low =0x02> <CRC high> <CRC low>
Response:
<Meter Address> <Command code=06> <Register address =0x00> <Register address =0x0a>
<Register data=0x00> <Register data =0x02> <CRC high> <CRC low>
Reset Total:
Address = 40011, data = 0x02
This command is used to clear the Totalizer and elapsed time registers
Reset 24 hours Total:
Address=40011, data = 180 (0xB4)
This command resets the 24 hours 7 days record including the day and hours counters
Reset 24 hour time:
Address=40011, data = 181 (0xB5)
This command resets the 24 hours day and hours counters.
24 hours Event:
Address=40011, data = 182 (0xB6)
This command generates a 24 hours event, the same way as when the 24 hours counter rolls over.
This may be useful to record total over a shorter period.
CAL-V Verify:
Address=40011, data = 161 (0xA1)
This command initiates a "CAL_V Verify". This operation may take 4 minutes to complete and
will stop the meter from calculating ow. The Status2 bit D0 may be monitored to check for
completion.
Zero Check In-Pipe Verify:
Address=40011, data = 173 (0xAD)
This command initiates a "Zero Check In-Pipe Verify". This operation does not aect ow
calculations. The register 40069 may be monitored to check for completion.