Chapter 4 Functional Checks 4-21
4-3-9-2 Doppler Mode OP Panel Controls
Table 4-7 Doppler Mode OP Panel Controls
Step Task Expected Result(s)
1
Press PW Mode key PW Mode Starts
2
Adjust Gain
Amplifies the overall strength of the echoes
processed in the Color Flow window. Turn the
Gain dial (PW Mode key) to the left/right to
increase/decrease Gain.
3
Display M/D-Mode Cursor
Displays the M/D-Mode cursor on the B-Mode
image. Press Cursor and Trackball to position
sample volume graphic. Click SV gate to
adjust sample volume gate size.
4
B-Pause
Toggle between simultaneous and update
presentation while viewing Spectral Doppler.
Press B Pause to toggle between
simultaneous and update.

Table of Contents