GE PROPRIETARY TO GE
D
IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-25
7-6-4-1 No Connection to the Network at All (cont’d)
2.) Check that the network cable between the scanner and the wall network is connected and well
seated in both ends.
3.) Try a network cable that is known to be OK.
4.) Connect a network cable between the Scanner and your PC. Try to ping from the scanner to the IP
address on the PC. If ok, the hardware connection inside the scanner is OK.