GE PROPRIETARY TO GE
D
IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
Chapter 10 Care & Maintenance 10-3
NOTE: May require specialized equipment to complete
NOTE: PMs are not mandatory, the table above is for reference only.
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Weekly Monthly
Per Facilities
QA Program Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Air filter •
More frequently depending
on your environment
Clean LCD
•
Inspect Wheels,Casters,brakes, and
Swivel locks
•
Mobile unit check
Daily
Console Leakage Current Checks
See Note
Twice
Annually
Peripheral Leakage Current Checks
See Note
Twice
Annually
Surface Probe Leakage Current Checks
See Note
Twice
Annually
Endocavity Probe Leakage
Current Checks
See Note
Quarterly
Annually
Measurement Accuracy Checks
See Note
Twice
Annually