IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
Chapter 10 Care & Maintenance 10-15
Leakage current limits (continued)
*ISO and Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to determine
the amount of current that will pass (or sink) to ground if a patient is in contact with mains voltage.
*Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table IV
Table 10-15: ISO and Mains Applied Limits*
Probe Type Measurement
BF 5.0 mA
CF 0.05 mA