DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
4 - 36 Section 4-9 - Board Function Checks
10.)Activate probe. Refer to the 4-3-7-2 "Activating the probe" on page 4-9.
11.)Check the system basic functions. Refer to the 4-3-4 "System B/M-Mode Checks" and refer to the
4-3-5 "System CFM and PWD Checks" on page 4-7.
12.)Disconnect probe. Refer to the 4-3-7-4 "Disconnecting the probe" on page 4-9.
13.)Connect probe on port 2 and repeat step 8) ~ 10).
14.)Connect probe on port 3 and repeat step 8) ~ 10).
15.)All basic functions must show correct operation as described.