DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
2 - 6 Section 2-2 - General Console Requirements
2-2-4 Scan Probe Environmental Requirements
Operation:10° to 30° C
Storage:-10° to 60° C
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
NOTICE
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
OF -10 TO + 60 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.