DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
4 - 46 Section 4-9 - Board Function Checks
4-9-11 DTRX Cooler with heatsink Function Validation Procedure
1.) Turn on the system
2.) Check the basic function of system's different modes. Refer to the 4-3-4 "System B/M-Mode
Checks" on page 4-6 and refer to the 4-3-5 "System CFM and PWD Checks" on page 4-7.
Refer to the 10-5-2-1 "System Checks" on page 10-8.
4-9-12 ACPC II Assy set Function Validation Procedure
1.) Turn on the system. Refer to the 3-5-1 "Power On/Boot Up" on page 3-11 sequence to check AP950
Assy function. The system must show correct operation as described on sequence.
2.) Check the peripheral operation. Refer to the Section 4-5 "Peripheral Checks" on page 4-21. All
peripheral must work correctly as described Section 4-5.