DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 5-1 - Overview 5 - 35
Section 5-8RFS (Service For Request)
5-8-1 General
This describes general RFS functionality.
1.) RFS function is a call for service to be used by customer or FE directly from the system instead of
calling GE Cares
5-8-2 System Checkout Procedure
If the system is checked out already, skip 5-8-2 - System Checkout Procedure and proceed from 5-8-
3 - Sending RFS Procedure.
1.) Check if the system is connected to the network.
2.) Go to “Utility”
3.) Press “Service”
4.) Log on to the Service Platform with a password.
5.) Press “Configuration”
6.) Press “InSite ExC Agent Configuration”
Figure 5-36 Agent Configuration