7 - 40 Section 7-9 - Trouble Shooting Tree
7-9-11 Network Troubleshooting
7-9-11-1 No Connection to the Network at All
1.) Check that the network cable between the scanner and the wall network is connected and well
seated in both ends.
2.) Try a network cable that is known to be OK.
3.) Check the cable between the SYSCONML assy and P6 ARP assy.
4.) Connect a Crossed network cable between the Scanner and your PC. Try to ping from the scanner
to the IP address on the PC. If OK, the hardware connection inside the scanner is OK.
7-9-11-2 No Contact With Server, But Connection to Network is OK
Check Network setup on the scanner.