8 - 138 Section 8-4 - Software Installation
8-4-4-9 Regional Setup Procedure for Japanese (cont’d)
3.) Select the Language tab, press Details, under Installed Services add the Japanese-Microsoft IME
Standard 2002 ver. 8.1 keyboard, under Default input language select Japanese, press Apply,
Press OK.
4.) Select the Advanced tab, then select Japanese in the Language for non-Unicode programs pull-
down menu. Press Apply. Answer Yes to use files already loaded on the hard disk, then answer No
to not reboot the system yet, press OK. Press Save and Exit the Utility screen.
5.) Reboot the system. When yur system restarts, the system appears in the selected language.
NOTE: To have the settings take effect, you MUST turn off the system and turn it back on.
6.) In Utility --> System --> General, press Regional Options. Select Language Tab and press Details.
Remove English under Installed Services. Press Apply and OK.
7.) To type in Japanese, press Alt + tilde/accent key (located between the Esc and F1 keys) in order to
start the Input Method Editor (IME).
Figure 8-208 Regional and Language Options—Language Tab

Table of Contents