Loading the software
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-35
5610739-100 English Rev.10
Wipe Tool
Wipe tool is intended to erase all the patient data with the
software to wipe the partitions on the system before the system
will be shipped for service.
NOTE: Wipe tool will be performed with the software USB flash drive.
Be sure your software USB flash drive is well kept.
NOTE: Before starting the wipe tool, please ensure that the power can
be continuously supplied and there is no risk of power cut off
during loading procedure.
1. Insert USB flash drive labeled “System & Application
Software” to the system.
2. Properly turn off the system by momentarily pressing the
Power/On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
NOTE: If the system will not shut down normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB flash
drive automatically.
CAUTION
The system can not be boot up after wipe process.
WARNING
The wipe procedure will erase all the patient data and all
existing software on the system. While the wipe procedure is
designed to preserved data, you should save any patient data,
images, system setups and customer presets to CD, DVD,
USB Flash Drive, or USB Hard Disk before doing the wipe
process.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals