4 Command and Query Summary
K0472 Revision A 4-96
:SOUR:SPE:LEV:IMM:AMPL
Applicability: PACE6000
Command Syntax
:SOURce[x]:SPEed[:LEVel][:IMMediate][:AMPLitude] <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> Airspeed in current units
Short form: SOUR:SPE <number>
Function: Sets the airspeed set-point
Default: 0.0
Query Syntax
:SOURce[x]:SPEed[:LEVel][:IMMediate][:AMPLitude]?
where x = 1 or 2 is the module number
(default - 1)
Short form :SOUR:SPE?
Function: Programmable set-point value
Response: Decimal number representing airspeed set-point in current units.
Description
This command sets the airspeed set-point. This query gets the decimal number representing
airspeed set-point in current units.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.