PACE Series SCPI Manual
K0472 Revision A 4-97
:SOUR:SPE:LEV:IMM:AMPL:SLEW
Applicability: PACE6000
Command Syntax
:SOURce[x]:SPEed[:LEVel][:IMMediate][:AMPLitude]:SLEW<number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> Airspeed rate set-point in current units per second.
Short form: SOUR:SPE:SLEW <number>
Function: Sets the rate set-point for airspeed.
Default: 0.0
Query Syntax
:SOURce[x]:SPEed[:LEVel][:IMMediate][:AMPLitude]:SLEW?
where x = 1 or 2 is the module number
(default - 1)
Short form :SOUR:SPE:SLEW?
Function: Queries the rate set-point for airspeed.
Response: Decimal number representing the rate set-point airspeed in currently
selected units of speed per second.
Description
This command sets the airspeed rate set-point. The query gets the rate set-point for
airspeed.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.