WP21 Series Operation Manual Chapter 5 Routine Operation
32
Temperature: 37 Sample volume: 500μL
Decimal places: 0
Unit: μ/L Blank: No Wavelength 1: 340
Sampling points: 2
Normal value (high): 40 Normal value (low): 0
Correction: 1.000, 0.00 Calibration: No Linear range: <500
Delay time: 30S Interval time: 40S Factor: 1746.0
2) Press the number key 1” on the “Item Choice” window to choose
the “ALT” item. Then, the system begins to preheat and will enter the
measurement interface after the temperature is stable. Enter the
number key “1” to select Measure”. When the system is ready, the
screen will display ASP Rinse”. Place the cleaning solution onto the
analyzer and press “1 Sample to start the washing operation. After
completion, the screen displays ASP DI Water”. Place the distilled
water onto the instrument and press “1 Sample to measure the
absorbance.
3) The screen displays ASP Sample”. Place the numbered patient
samples onto the instrument in order and press “1” to start
measurement of sample.
4) After aspirating the sample solution, the analyzer will display the
reaction curve on its screen during testing, as illustrated in Figure 5-21: