WP21 Series Operation Manual Chapter 5 Routine Operation
33
Figure 5-21
II. Determination of total protein by end-point method
1) Set the following parameters of this item in the “Program” menu
according to the reagent instruction:
Name: TP Method: End-point method
Number of wavelengths: 1
Temperature: 37℃ Sample volume: 800μL Decimal places: 1
Unit: μ/L Blank: Reagent Wavelength 1: 546
Sampling points: 1
Normal value (high): 82.0 Normal value (low): 62.0
Correction: 1.000, 0.00 Calibration: No Linear range: <100.0
Delay time: 3S Standard: 70.0
2) Press the number key “5” on the “Item Choice” window to choose
the “TP” item. Then, the system begins to preheat and will enter the
measurement interface after the temperature is stable. Enter the
number key “1” to select “Measure”. When the system is ready, the