WP21 Series Operation Manual Chapter 5 Routine Operation
34
screen will display ASP Rinse”. Place the cleaning solution onto the
analyzer and press the key “1” to start the washing operation.
3) After completion, the screen displays “ASP DI Water”. Place distilled
water onto the instrument and press “1” to measure the absorbance.
After testing, the screening display ASP Blank Solution”. Place the
blank solution onto the analyzer and press “1” to confirm the operation.
4) At this time, the screen displays ASP Standard Solution” (if the
screen displays ASP Sample”, please press the number key “2” to
open the ASP Standard Solution” window). Place the standard
solution onto the instrument and press “1” to perform measurement.
Calculate the factor after the testing is completed.
5) After measurement of the standard solution, the screen displays
ASP Sample”. Place the numbered patient samples onto the
instrument in order and press Sample to start measurement of
sample solutions.