USER HANDBOOK
SAFETY
The instrument must be regularly serviced and calibrated by fully trained
personnel in a safe area thus user adjustment is not required.
Replacement or charging of the equipment is only permitted in a non-
hazardous area.
Only GMI replacement parts should be used.
Instrument must be re-charged in a safe area before use. Refer to
Chapter 6 : ‘RECHARGE BATTERY’.
If the instrument detects gas, follow your own organisation’s procedures
and operational guidelines.
Gas can be dangerous and care should always be taken in its use.
This equipment is designed and manufactured to protect against other
hazards as dened in ATEX Directive 94/9/EC.
Any right of claim relating to product liability or consequential damage to any third
party against GMI is removed if the above warnings are not observed.
ADDITIONAL SAFETY DETAILS - CSA ONLY
CAUTION: Before each days Usage, test on a known concentration of methane,
equivalent to 25 - 50% of full scale concentration. Accuracy must be within
0 to +20% of actual. Accuracy may be corrected by calibration. (Refer to
Chapter 7 : ‘CALIBRATION’).
ATTENTION: Avant chaque utilisation journalière, testez la réaction de l’appareil
en utilisant une concentration connue en méthane, correspondant à 25-50%
de la lecture à fond d’échelle. La précision doit se situer entre 0 et +20% de la
valeur réelle. La précision peut se corriger en étalonnant l’appareil (voir Chapitre
7 : CALIBRATION).
CAUTION: Any rapid up-scale readings followed by a declining or erratic reading
may indicate a gas concentration beyond the upper scale limit, which may be
hazardous.
ATTENTION: Toute lecture dépassant rapidement le maximum de l’échelle et
suivie par une diminution ou une lecture erronée, indique une concentration de
gaz supérieure à la valeur maximale de cette échelle. Cette lecture n’est pas
signicative.
CAUTION: Substitution of components may impair intrinsic safety.
ATTENTION: le remplacement d’un composant peut porter atteinte à la sécurité
intrinsèque du produit.
CAUTION: Do not charge in a hazardous area. Um = 6V.
ATTENTION: Ne pas charger en zone dangereuse. Um = 6V.
NOTE: CSA have only assessed the LEL combustible gas detection portion of
this instrument for performance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GMI PS241 and is the answer not in the manual?

GMI PS241 Specifications

General IconGeneral
TypePressure Switch
Protection ClassIP65
Power Supply24 VDC
Contact TypeSPDT
Process ConnectionG1/4"
Operating Temperature-20...80°C

Related product manuals