4-1
ALARMS
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE
AN EXPLOSIVE CONCENTRATION.
4.1 GAS ALARMS
Gas alarms are enabled when the instrument is switched on
and warm-up is complete.
All gas ranges have alarm limits that trigger the alarm if the
measured gas value exceeds the set level. If a preset alarm
level is exceeded, the instrument vibrates, the display backlight
illuminates red, the audible alarm sounds, the LED’s ash red
and the gas range in alarm ashes on the display.
The alarms are individually programmable to be either
‘Latching’ or ‘Non-Latching’.
A ‘Latching’ alarm will stay on until reset by the user with a
press and hold of the Right Hand (RH) button when
the gas readings are safe.
A ‘Non-Latching’ alarm will reset automatically when the gas
readings are safe.
The following table illustrates the factory default alarm
indications: