The bump test result including date and time will be
automatically datalogged.
When the gas readings fall below their alarm set-points, or after
60 seconds, the bump test is complete, and the instrument will
automatically return to normal operation.
Should any gas range fail the bump test, the alarm conrmation
display will be red and a symbol will be displayed as shown
in Fig. 3-7.
Fig. 3-7 Bump Test Fail - Alarm Conrmation
After selecting ‘YES’ or ‘NO’ the user is informed of the bump
test fail and instructed to switch the instrument off, as shown
in Fig. 3-8.
Fig. 3-8 Bump Test Fail - Switch Off PS200
If an instrument fails a bump test, then a full re-calibration is
necessary.