3-6
USER HANDBOOK
The bump test result including date and time will be
automatically datalogged.
When the gas readings fall below their alarm set-points, or after
60 seconds, the bump test is complete, and the instrument will
automatically return to normal operation.
Should any gas range fail the bump test, the alarm conrmation
display will be red and a symbol will be displayed as shown
in Fig. 3-7.
Fig. 3-7 Bump Test Fail - Alarm Conrmation
After selecting ‘YES’ or ‘NO’ the user is informed of the bump
test fail and instructed to switch the instrument off, as shown
in Fig. 3-8.
Fig. 3-8 Bump Test Fail - Switch Off PS200
If an instrument fails a bump test, then a full re-calibration is
necessary.