3-2
USER HANDBOOK
3.2 MANUAL BUMP OPTIONS
The PS200 series provides two bump test options, QUICK
and FULL.
The QUICK bump test validates that the alarm threshold has
been exceeded for each range.
The FULL bump test checks the response of all ranges against
set limits.
By default, both of these options are disabled.
3.3 INITIATING A MANUAL BUMP TEST
To initiate a manual bump test, switch the instrument ON with
a long press of the LH button.
During warm-up, the user will be prompted to conrm a bump
test is required, as indicated in Fig. 3-1. Press the RH button
to continue.
Fig. 3-1 Conrm Bump Test
When warm-up is complete, the user is prompted to apply test
gas, as shown in Fig. 3-2.