4-6
USER HANDBOOK
If the LEL sensor is exposed to a gas reading above 100%
LEL, the instrument vibrates, the displayed value changes
to 4 rising arrows, the backlight illuminates red, the tone of
the audible alarm changes, and the visual alarm LED’s ash
quickly. The ashing message ‘DANGER OVER RANGE’ is
displayed, as shown in Fig 4-3:
Fig. 4-3 Over Range Alarm
Switch OFF by a press and hold of both buttons together.
A timer, counting down from 10 seconds to zero, is displayed
together with the message ‘GET OUT’ alternating with ‘HIGH
GAS’, as shown in Fig. 4-4:
. . . alternating until zero is reached
Fig. 4-4 ‘Get Out’ / ‘High Gas’ Timer
The instrument must be returned to a gas free area or sample
clean air. The instrument must now be switched OFF.