5-5
OPERATOR MAINTENANCE
Fig. 5.3 In-line Hydrophobic Filter
To replace the lter, proceed as follows:
1. Unscrew the luer tting from one side of the lter in a
counter clockwise direction, detach the tubing from the
other side then remove the hydrophobic lter.
Note: If re-tting the same lter, make sure that lter
direction of ow orientation is maintained. This can be
easily identied by position of yellow label on lter, i.e.
facing instrument.
2. Fit a new In-line Hydrophobic Filter (Part No. 66484).
Note that the lter should be tted with the yellow label
facing the instrument.
3. Attach the luer tting, with sample line attached, to
the mating side of the lter then tighten in a clockwise
direction to secure. Do not overtighten the tting.
4. Attach the sample line to the other side of the lter making
sure that it is securely tted.