80 GMC-I Messtechnik GmbH
12 Characteristic Values
Func-
tion
Measured
Quantity
Display Range/
Nominal Range of
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nomi-
nal
Current
I
N
Short-
Circuit
Current
I
K
Internal
Resis-
tance
R
I
Refer-
ence
Resis-
tance
R
REF
Measuring
Uncertainty
Intrinsic
Uncertainty
Overload
Capacity
Value Time
Tests, 62638 (DIN VDE 0701-0702) / IEC 62353 (VDE 0751)
Protective conduc-
tor resistance
R
PE
1 999 mΩ 1mΩ
< 24 V
AC or DC
>
200 mA
AC / DC
> 10 A AC
5
>
35 A AC
11
——
±(15% rdg.+ 10 d)
>10d
> 10.0 Ω :
±(10% rdg.+ 10 d)
±
(10% rdg.+ 10 d)
>10d
264 V
250 mA
16 A AC
5
Continu-
ous
1.00 ... 9.99 Ω 10 mΩ
10.0 27.0 Ω
100 m
Ω
>
42 A AC
11
15 s
Insulation resis-
tance
9
RINs
10 999 kΩ 1kΩ
50 500
VDC
1.0 • U
N
1.5 • U
N
>1mA <2mA
±(5% rdg.+ 4 d)
>10d
±
(2.5 % rdg.+2 d)
> 10 d
264 V
Continu-
ous
1.00 9.99 MΩ 10 kΩ
10.0 99.9 MΩ 100 kΩ
20 MΩ :
±(10% rdg.+ 8 d)
20 MΩ :
±
(5% rdg.+4 d)
100 300 MΩ 1MΩ
Leakage current
Alternative
Measurement
2
IPE, IT, IE, IA
0 ... 99 μA1μA
50 ...
250 V~
20/+10%
—<1.5mA
> 150 k
Ω
1kΩ
±10 Ω
±
(5% rdg.+ 4 d) > 10 d
> 15 mA:
±(10% rdg.+ 8 d)
±
(2% rdg.+2 d) > 10 d
> 15 mA:
±
(5% rdg.+ 4 d)
264 V
Continu-
ous
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current
Direct measure-
ment
3
IPE, IT, IE, IA, IP
Ip only: 0.0 ...
99.9 μA
100 nA
——
1kΩ
±10 Ω
1kΩ
±(5% rdg.+ 4 d)
> 10 d
±
(2.5 % rdg.+2 d)
> 10 d
264 V
Continu-
ous
0 ... 99 μA1μA
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current
Differential cur-
rent measure-
ment
4
IPE, IT, IE
0 ... 99 μA1μA
————
±(5% rdg.+ 4 d)
> 10 d
±
(2.5 % rdg.+2 d)
>10d
264 V
Continu-
ous
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Function test at the test socket
Line voltage
U
L–N
10
100.0 240.0 V~ 0.1V ————
±
(2% rdg.+2 d)
264 V
Continu-
ous
Load current I
L
0 16.00 A
RMS
10mA ————
±
(2% rdg.+2 d)
16 A
Continu-
ous
Active power P
0 3700 W 1 W
±
(5 % rdg.+10 d)
>20d
264 V
Continu-
ous
20 A 10 min.
Apparent power S
0 4000 VA 1 VA Calculated value, U
L–N
• I
V
±
(5 % rdg.+10 d)
>20d
264 V
Continu-
ous
Power factor PF
with sinusoidal
waveform: cosϕ
0.00 1.00 0.01 Calculated value, P / S, display > 10 W ±(10% rdg.+5 d) 264 V
Continu-
ous
Line frequency f
0 ... 420.0 Hz 0.1 Hz
±
(2% rdg.+2 d)
264 V
Continu-
ous
t
A
PRCD
Time to Trip
0.1 ... 999 ms 0.1 ms 30 mA
±
5 ms 264 V
Continu-
ous
Voltage measurement
Probe voltage
(probe P1 to PE)
, and
0.0 ... 99.9 V
100 ... 264 V
100 mV
1 V
——
3 MΩ
——
±
(2% rdg.+2 d)
264 V
Continu-
ous
Measuring voltage
(V–COM
sockets
6
)
, and
0.0 ... 99.9 V
100 ... 300 V
1 MΩ
±
(2 % rdg.+2 d)
> 45 Hz ... 65 Hz
±
(2 % rdg.+5 d)
> 65 Hz ... 10 kHz
±
(5 % rdg.+5 d)
> 10 kHz ... 20 kHz
300 V
,
and
I
Leaka
ge
Leakage current
via AT3-IIIE
adapter Z745S
6, 8
0.00 ... 0.99 mA
0.01 mA
————
±(2 % rdg.+2 d)
>10d
without adapter
253 V
Continu-
ous
1.0 ... 9.9 mA
0.1 mA
10 ... 20 mA
1mA
Temp
Temperature
with Pt100 sensor
200.0 ...
+850.0 °C
0.1 °C— <20V 1.1mA ±(2% rdg.+1 °C) 10 V
Continu-
ous
Temperature
with Pt1000 sen-
sor
150.0 ...
+850.0 °C

Table of Contents

Other manuals for Gossen MetraWatt SECUTEST PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gossen MetraWatt SECUTEST PRO and is the answer not in the manual?

Gossen MetraWatt SECUTEST PRO Specifications

General IconGeneral
BrandGossen MetraWatt
ModelSECUTEST PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals