GPT-10000 Series User Manual
124
Return to
READY
status
Restore to READY status
Note
When in FAIL status, only the STOP button can be
pressed, all other keys are disabled.
Stop a
Running Test
1. To stop the AUTO test at any time
when it is running, press the STOP
button. The AUTO test will stop
immediately. When the STOP
button is pressed, a judgment is
not made on the current test and
any remaining tests are aborted.
All panel keys except the STOP
and START buttons are disabled
when the tester has been stopped.
All the results up until when the
AUTO test was stopped are shown
on-screen. See page 126 for more
details on automatic test results.
Below is example of an automatic test
that has been stopped in the midway.
The remaining MANU STEPs are
aborted without test results.