REMOTE CONTROL 
235 
Parameter/ 
Return parameter 
MANU:DCW:INIT 87 
Sets the DCW Initial Voltage to 87%. 
 
Sets or returns the CONTACT CHK function on or off. 
Note: An “Err” message will be shown on display if 
the learned reference value is less than 30uA. An 
“LEARNING < 30uA” message will be returned after 
using the query command “SYSTem:ERRor ?” in 
remote control. 
MANU:DCW:CONTact {ON|OFF} 
MANU:DCW:CONTact? 
Parameter/ 
Return parameter 
CONTACT CHK in DCW test ON 
CONTACT CHK in DCW test OFF 
MANU:DCW:CONT OFF 
Sets the CONTACT CHK off in DCW test. 
 
In certain applications, it is suggested to enable 
Filter On, which indicates the activation of 
software judgment function that judges values by 
interval of every 100ms in case of instantaneous 
misjudgment on distorted waveforms.   
MANU:DCW:FILTer {OFF|ON} 
MANU:DCW:FILTer ?