GPT-10000 Series User Manual 
250 
 
Sets or returns the CONT reference value in Ω. The 
test must first be in CONT mode before this 
command can be used. 
Note: when HI SET+REF VALUE is over 80 Ω, an 
“Err” message will be shown on display. And an 
“CONT Setting Over 8V” message will be returned 
after using the query command “SYSTem:ERRor ?” 
in remote control. 
MANU:CONTinuity:REF <NR2> 
MANU:CONTinuity:REF? 
Parameter/ 
Return parameter 
MANU:CONT:REF 0.01 
Sets the CONT reference to 00.01 Ω. 
MANU:CONTinuity:ZEROCHECK 
 
Performs the zero check function.  The test must 
first be in CONT mode and in the Ready Status 
before this command can be used. 
 
See page 71 for details on the ZERO function. 
MANU:CONTinuity:ZEROCHECK {ON|OFF} 
MANU:CONTinuity:ZEROCHECK? 
Parameter/ 
Return parameter 
Zero function is not active. 
MANU:CONT:ZEROCHECK OFF 
Deactivates the CONT ZERO function.