REMOTE CONTROL
237
MANU:DCW:LOWLimit
Set
Query
Description
Sets or returns the threshold of Low Limit scale for
contact check function in the specific GPT-15012 unit.
Syntax
Query Syntax
MANU:DCW:LOWLimit {value}
MANU:DCW:LOWLimit ?
Parameter
<value>
5% ~ 95%
Return parameter
<value>
5% ~ 95%
Example
MANU:DCW:LOWL 80%
Sets the threshold of Low Limit scale as 80% to the
reference value.
MANU:IR:VOLTage
Set
Query
Description
Sets or returns the IR voltage in kV. The test must
first be in IR mode before this command can be
used.
Syntax
Query Syntax
MANU:IR:VOLTage <NR2>
MANU:IR:VOLTage?
Parameter/
Return parameter
<NR2>
0.05 ~ 5 (0.05kV to 5kV: steps of .05)
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value in
M or GΩ. The test must first be in IR mode before
this command can be used.
Syntax
Query Syntax
MANU:IR:RHISet <NR2>|NULL
MANU:IR:RHISet?

Table of Contents

Other manuals for GW Instek GPT-12004

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-12004 and is the answer not in the manual?

GW Instek GPT-12004 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-12004
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals