GPT-10000 Series User Manual
214
MEAS?
> CON,FAIL ,100.0mA,99.99 ohm,T=000.1s
Returns the test result of the current MANU test.
MEAS21?
> DCW,FAIL ,0.004kV, 000.0 uA ,T=000.3s
Returns the step 21 of the current AUTO test result.
Changes the mode between AUTO and MANU.
MAIN:FUNCtion {MANU|AUTO}
MAIN:FUNCtion ?
Parameter/
Return parameter
Puts the tester mode to MANU.
Puts the tester mode to AUTO.
MAIN:FUNC MANU
Sets the tester to MANU mode.
Turns on or off the “OK" message for test result,
which is shown when a test finishes.
TESTok:RETurn {ON|OFF}
TESTok:RETurn ?
Parameter/
Return parameter
Turns on the “OK" message for test result.
Turns off the “OK" message for test result.
TEST:RETURN ON
Turns of the OK message return function.