REMOTE CONTROL
243
Example
MANU:IR:FILT OFF
Sets the FILTER off in IR test.
MANU:IR:GNDoffset
Set
Query
Description
Sets or returns the GND Offset function for IR test.
Syntax
Query Syntax
MANU:IR:GNDoffset {ON | OFF}
MANU:IR:GNDoffset?
Parameter/
Return parameter
ON
GDN Offest in IR test On
OFF
GDN Offest in IR test Off
Example
MANU:IR:GND OFF
Sets the GND Offest off in IR test.
MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.
Syntax
Query Syntax
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/
Return parameter
<NR2>
3.00~33.00
Example
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.
MANU:GB:RHISet
Set
Query
Description
Sets or returns the GB HI SET resistance value in
m. The test must first be in GB mode before this
command can be used.
Syntax
Query Syntax
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/
Return parameter
<NR2>
000.1 ~ 650.0 (m)

Table of Contents