EasyManua.ls Logo

GW Instek GPT-12004 - System:cal:protection; System:internal:memory

GW Instek GPT-12004
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
209
Example
SYST:ALERT:DATE T21_06_01
Sets the pre-alert date as 2021-06-01.
SYSTem:CAL:PROTection
Set
Query
Description
Sets or returns the calibration output protection for
calibration due date.
Syntax
Query Syntax
SYSTem:CAL:PROTection {ON |OFF}
SYSTem:CAL:PROTection?
Parameter
ON
Turns the calibration protection on.
OFF
Turns the calibration protection off.
Return parameter
ON
Calibration protection on.
OFF
Calibration protection off.
Example
SYST:CAL:PROT ON
Sets the calibration protection function ON.
SYSTem:INTernal:MEMory
Set
Query
Description
Returns the test result from internal memory of unit.
Query Syntax
SYSTem:INTernal:MEMory?
Return parameter
<string>
Returns the test result from
internal memory as a string.
Example
SYST:INTERNAL:MEMORY?
> Date/Time ,Stp,MOD,Judge, V/A , Reading ,
Timer ,MA_S,Barcode ,
2021-02-24 13:52:08,001,IR ,PASS ,0.049kV,
>10Gohm,T=000.3s,A002, ,

Table of Contents

Other manuals for GW Instek GPT-12004

Related product manuals