GPT-9000/9000A Series User Manual
102
7. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
Ensure the baud rate settings or GPIB address
matches the host machine.
Note
The ESC key can be pressed at any time to cancel
and exit back to VIEW status.
Control Settings
Description
The Control settings are accessed in the
COMMON UTILITY menu. The Control
settings include: Start Control, Double Action,
Key Lock and Interlock.
Start Control is used to determine how a test is
started. Tests can be started via the front panel
(START/STOP buttons), from a remote
controller or via the SIGNAL I/O port.
The Double Action function is a safety feature
used to prevent accidentally starting a test.
Normally to start a test, the START button is
pressed when the tester is in the READY status.
To start a test when Double Action is ON, the
STOP button must first be pressed, followed by
the START button within 500ms.
Key Lock disables the front panel keys from
changing the test number, mode or testing
parameters. Only the Utility menu and any
keys required for testing are not disabled.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals