CONFIGURATION
103
Internal
Resistance
Settings
Sets the internal resistance of the power supply.
F-08
0.000~ X.XXXΩ
(Where X.XXX = Rating Voltage /
Rating Current)
Bleeder Control
Bleeder control turns ON/OFF the bleeder resistor.
Bleeder resistors discharge the filter capacitors
after power is turned off as a safety measure.
F-09
0 = OFF, 1 = ON, 2 = AUTO
Buzzer ON/OFF
Turns the buzzer sound on or off. The buzzer is
associated with alarm sounds and keypad entry
sounds.
F-10
0 = OFF, 1 = ON
OCP Delay Time
Sets the OCP delay time. This parameter will delay
the amount of time it takes to trigger the over
current protection. This function can be useful to
prevent current overshoot from triggering OCP.
F-12
0.1 ~ 2.0 sec
Current Setting
Limit (I-limit)
Turns the current setting limit (I-limit) on or off.
Turning this function on will prevent you from
accidentally setting the current limit above the set
OCP level.
F-13
0 = OFF, 1 = ON
Voltage Setting
Limit
Turns the voltage setting limit (V-limit) on or off.
Turning this function on will prevent you from
accidentally setting the voltage limit above the
OVP level.
F-14
0 = OFF 1 = ON
Display Memory
Parameter
Displays which memory setting is recalled (M1,
M2 or M3) when recalling a setup.
F-15
0 = OFF, 1 = ON

Table of Contents

Other manuals for GW Instek PSU 100-15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU 100-15 and is the answer not in the manual?

GW Instek PSU 100-15 Specifications

General IconGeneral
BrandGW Instek
ModelPSU 100-15
CategoryPower Supply
LanguageEnglish

Related product manuals