ANALOG CONTROL
151
Connection- alt.
shielding
If the wire shield needs to be grounded at the
current source (EXT-I), then it can be connected as
shown below.
PSUEXT-I
Isolated
Current
Option
4
2
Output
Terminal
2 core shielded
wire or twisted
pair
Pin4(GND_ISO) → EXT-I (-)
Pin2(+VPROG_ISO) → EXT-I (+)
EXT-I ground (GND) → Wire shield
Panel operation
1. Connect the external current source according
to the connection diagrams above.
2. Set the F-90 power on
configuration setting to 4 (Control
by Isolated Board).
Page 114
3. Set the F-96 power on configuration setting to 1.
4. Set the F-97 power on configuration setting to 1.
Be sure to cycle the power after the power on
configuration has been set.
5. Press the Function key and confirm
the new configuration settings
(F-90=4, F-96=1, F-97=1).
Function
6. Press the Output key. The
voltage can now be
controlled with the isolated
external current control.
Output

Table of Contents

Other manuals for GW Instek PSU 100-15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU 100-15 and is the answer not in the manual?

GW Instek PSU 100-15 Specifications

General IconGeneral
BrandGW Instek
ModelPSU 100-15
CategoryPower Supply
LanguageEnglish

Related product manuals