EasyManuals Logo

GW Instek PSU 100-15 User Manual

GW Instek PSU 100-15
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
PSU Series User Manual
80
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) should be performed on
both units.
The VMON voltage represents the voltage of
that particular unit.
Remote Sense
Please see the voltage sense chapter for details,
page 61.
External Voltage and Resistance Control
Voltage/Resistance controlled remote control
should be used on both units separately.
The full scale voltage (in series) is equivalent to
the maximum external voltage or resistance.
Slew Rate
The slave rate should be set for both units.
Internal Resistance
The internal resistance should be set for both
units.
Bleeder Control
The bleeder resistor setting should be set
equally on both units.
When using analog control connector to program
or measure with PSU power supplies connected in
series, make sure that each unit is separated and
floating from each other.
When PSU power supplies are connected in series
and the load or one of the output terminals is
grounded, no point on the output shall be more
than ±600VDC above or below chassis ground.

Table of Contents

Other manuals for GW Instek PSU 100-15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU 100-15 and is the answer not in the manual?

GW Instek PSU 100-15 Specifications

General IconGeneral
BrandGW Instek
ModelPSU 100-15
CategoryPower Supply
LanguageEnglish

Related product manuals