Electrical installation
Wiring safety information
D A N G E R
Electrocution hazard. Always remove power to the instrument before making electrical connections.
About quick-connect wiring connections
D A N G E R
Electrocution and fire hazards. For instruments with a power cord, make sure that there is easy access
to the local power disconnect.
Figure 5 shows the quick-connect connections.
Figure 5 Power supply and wire connections
1 Power switch 5 RS232 connections
2 Power cord connection 6 Network connection
3 Digital output 7 Digital input
4 USB port 8 4–20 mA output connection
Connect the power cord
1. Connect the power cord to the instrument. Refer to Figure 5 on page
13, item 2.
2. Tighten the compression fitting by hand.
Connect the 4–20 mA output
Table 2 shows the terminal designations for the 4–20 mA quick-connect connector. The default
parameter represented on each 4–20 mA output is shown in Table 2. To change the default
parameter, refer to Configure the analog outputs on page 33.
English 13