4. Select the Network tab.
5. Select an option.
Option Description
Static TCP/IP
Address
Set the TCP/IP address for the analyzer. Options: 0 to 255 for each octet (default =
192.168.0.105)
Static Subnet Mask Set the subnet mask for the analyzer. Options: 0 to 255 for each octet (default =
255.255.255.0)
Static Gateway Set the default gateway for the analyzer. Options: 0 to 255 for each octet (default =
blank)
DNS Server Set the DNS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
WINS Server Set the WINS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
DHCP Enable DHCP. When selected, the TCP/IP address, subnet mask and default
gateway entries are disabled (default = enabled).
Configure the backlight settings
1. Push << to open the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
4. Select the Backlight tab.
5. Configure the backlight options.
Option Description
Automatically turn off backlight Have the backlight switch off automatically after a period of inactivity.
Turn off after Select the number of minutes of inactivity before the backlight switches
off. Options: 2, 5, 10, 15 or 30 minutes
6. To adjust the brightness of the backlight, push the brightness slider.
Configure the digital output alarms
Assign output functionally for each digital output as necessary. Refer to Table 13. To invert the digital
outputs, refer to Configure digital control on page 28
.
Note: The TOC alarm can be assigned to multiple digital outputs (1–4) at the same time to set a multi-level alarm
on the TOC result.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push ALARMS.
4. Assign the output functionality for each digital output. Refer to Table 13.
Note: Conductivity alarm, Error state and Valve state can only be assigned to one digital output at a time.
5. Select the checkbox to enable the alarm, if present.
6. For TOC alarm, select the upper TOC limit for the alarm. Options: 1 to 2500 (default = 500). This
value applies only to online TOC analysis, not to bottle or grab sample analysis.
English 31