3. Select all of the data types that apply. All types are selected by default.
4. Select time range or special range. Special range options sets the data to sort first by type, then
by time. The time range option sets the data to sort based on the specified date range.
Option Description
Last accepted Includes only the last accepted tests for the selected data types
Last done Includes only the last completed tests for the selected data types
Last five Includes only the last five completed tests for the selected data types
Export or print data
1. Push << to show the sliding menu.
2. Select DATA REVIEW.
3. Select an option.
Push EXPORT to send data as formatted text to the supplied USB flash drive.
Push PRINT to send data to an attached printer.
Advanced operation
TOC operating mode configuration
Configure the general settings
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the General tab.
5. Select an option.
Option Description
Flush Time Set the amount of time the cell is flushed before a TOC analysis is started (hh:mm:ss).
Options: 00:00:00 to 23:59:59 (default = 00:01:00)
Cycle Time Set the minimum amount of time between automatic TOC analyses. Any value less than
the actual elapsed time causes the next analysis to start immediately (hh:mm:ss).
Options: 00:00:00 to 99:59:59 (default = 00:00:00)
Use pump Enable the sample pump for online sampling (default = disabled). Do not enable the
sample pump on a pressurized system or pump damage may occur.
Bypass flow
during analysis
Enable bypass flow. When disabled, the bypass flow valve stays closed. When enabled,
sample water flows continuously through the analyzer (bypasses the analysis cell) during
TOC analysis. The bypass valve is open when the sample is captured and closed when
the analysis is complete.
Average last Average the last x number of TOC analysis. The averaged value is shown on the home
screen and in the log view, logged to the data log and sent to the printer based on the
printer settings.
Switch Lamps
when Marginal
Enable the analyzer to switch to the other UV lamp when the UV lamp diagnostics sense
a marginal level of UV intensity on the operating UV lamp.
English 27