Bottle mode configuration
Configure the TOC calibration settings
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push BOTTLES.
4. Select the TOC Calibration tab.
5. Select an option.
Option Description
User must manually
accept results after
test is complete
The results of the TOC calibration are not used until they are manually accepted.
Return to online mode The analyzer goes back to the online mode after the TOC calibration. When not
selected, the analyzer stays at the bottle mode dialog box. All valves are closed.
The pump is off. The lamp is off.
Note: This option is only available when "User must accept results after test is
complete" is selected.
Auto accept results
and return online
The Slope Limit and R2 Limit options that follow are enabled. If the calibration is
accepted, the new calibration factors are used.
Slope Limit <= Set the limit of the slope for the calibration to pass. Options: 1 to 15% (default =
15 %)
Note: This option is only available when "Auto accept results and return online" is
selected.
R2 Limit >= Set the limit for the determination coefficient. Options: 0.9000 to 0.9999 (default =
0.9900)
Note: This option is only available when "Auto accept results and return online" is
selected.
Display 'as Received'
reports
Enable 'as Received' reports to be shown in the calibration results.
Configure the conductivity calibration settings
1. Push <<
to show the sliding menu.
2. Push the SETUP icon.
3. Push BOTTLES.
4. Select the Conductivity Calibration tab.
5. Select an option.
Option Description
User must manually
accept results after test
is complete
The results of the conductivity calibration are not used until they are manually
accepted.
Return to online mode The analyzer goes back to the online mode after the conductivity calibration.
When not selected, the analyzer stays at the bottle mode dialog box. All valves
are closed. The pump is off. The lamp is off.
Note: This option is only available when "User must accept results after test is
complete" is selected.
Auto accept results and
return online
The Slope Limit options that come next are enabled. If the calibration is
accepted, the new calibration factors are used.
English 37